6-8 z.one ultra and z.one ultra sp OPERATOR MANUAL
WARNING: Any visible evidence of damage or improper mechanical operation
indicates the transducer cannot be used and should be returned to
ZONARE for
evaluation and repair.
See page 15-1 for contact information.
General Safety NOTE: See also the Safety Manual.
The multiplane TEE transducer should be used only by a qualified physician
who has received appropriate training in proper operation of the transducer
and in endoscopic techniques as dictated by current relevant medical
practices.
The multiplane TEE transducer is a precision instrument that must be handled
with care. It may be damaged when dropped or abused. In particular, do not
allow the acoustic window in the tip of the gastroscope to come into contact
with a sharp object. Do not touch this window unnecessarily. Never exert force
onto the acoustic window.
The transducer connector is not watertight, and should always be kept dry. The
control handle, although spray-watertight, should not be immersed.
This equipment contains no operator serviceable components. To prevent
electric shock, do not remove any covers or panels.
Avoid forceful manipulations and excessive force in using the transducer that
could result in patient injury.
Withdraw the transducer only with the deflection control in the unlock mode
and with the distal end straight.
Never manually deflect the distal tip; use only the deflection control wheels.
Do not kink, tightly coil, or apply excessive force on the transducer cable or
shaft. Insulation failure may result.
Under normal conditions at full acoustic power the temperature of the tip does
not exceed 43°C. Be sure to check at least monthly that the temperature
increase of the tip is within limits:
Connect the transducer to the ultrasound system.
Adjust the acoustic power to the highest value possible.
Select Color Doppler mode.
Wait for 2 minutes.
Feel at the distal end of the transducer if there is a temperature increase
that could be harmful for the patient.
Preparation for
Use
Clean and disinfect the transducer described in “Cleaning & Disinfecting TEE
Transducer” on page 6-15.
Inspect the transducer as described in “Inspecting Transducer” on page 6-14.
Connect the transducer to the transducer port on the side of the ultrasound
machine. The transducer may need to be activated by the Transducer button
on the Control Panel.
For patient protection a sterile, single-use, latex sheath can be used.
CAUTION: Ensure that the sheath's sterile surface is maintained.