7
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-75
NOTE: Once a Mode key is defined, it will remain so unless the system
configuration is reset to factory default settings.
2. Connect and activate a transducer enabled for Contrast Imaging: P4-1c
(Cardiac) or C4-1 (Abdominal).
3. Select an appropriate Preset for the application.
4. While imaging the desired anatomy, press Mode/Function key you
programmed for Contrast.
•For P4-1c, system enters Contrast mode & displays single image.
•For C4-1, system enters Simul Dual mode with Contrast detection image on
left and non-contrast reference image on right.
5. Once in Contrast mode, system displays Contrast softkey controls:
Figure 7-40. Contrast Imaging Softkeys
NOTE: Rotate Gain softkey to control Contrast image brightness. Rotate 2D/B
button to control reference image brightness.
Contrast Imaging User Controls
Contrast imaging controls are accessed via the Control Panel (all systems),
softkey displays (
SmartCart/SmartCart sp), and Imaging menu options (all
systems). They are described in
Table 7-8.
Compound
ON/OFF
Gain
Simul
[Frame
Rate Hz]
[Power
%]
Stopwatch
START
[STOP]
Stopwatch
RESET
[PHASE]
Table 7-8. Contrast Imaging User Controls
Control SmartCart & SmartCart sp
Softkeys
Simul Softkey (C4-1
transducer only)
Controls whether Dual Simultaneous or Dual (non-simultaneous) mode is
operational. By default, Dual Simultaneous mode is enabled.
■ Press Simul to transition to Dual mode, displaying only the Contrast detection
image.
■ Press Enter button to toggle between right and left imaging panes.
■ Press Simul again to return to Dual Simultaneous mode.