z.one ultra and z.one ultra sp OPERATOR MANUAL 9-15
To Export Data
1. Make sure the targeted Export system is connected to one of the external
USB ports of the System.
2. Perform the desired calculations and measurements on the scan.
3. To Export the data, select Export at the bottom of the Report page.
NOTE: If the Export was configured for Print/Export on End of Exam, data will be
automatically exported at the end of the current exam.
Generic Measurement Procedures
B-Mode
Depth Depth measures the distance from the transducer face to the location of the
caliper in the image. The Depth measurement result is displayed in the DDB and
updated continuously as the user positions the caliper within the image.
To Measure Depth
Live Image
1. Display a live B-Mode image.
2. Press Measure button.
3. Use Trackball to position caliper to desired location within image display. A
Depth result displays in DDB and updates as user positions caliper within the
image.
4. Press Measure button to remove single caliper and exit Generic
Measurements.
Frozen Image
1. Freeze a B-Mode image.
2. Press Measure button.
3. Use Trackball to highlight and select Distance. Then proceed to Steps 3 and 4
above.
Distance The Distance measurement displays the span between the two B-Mode calipers
in the image.
Up to four (4) Distance measurements can be displayed on a single image
simultaneously.
•The Distance measurement results are displayed in the DDB.
The active measurement result updates continuously as the user positions
the active caliper within the image.
To Measure Distance
1. Freeze the B-Mode display.
2. Press the Measure button.