14-20 z.one ultra and z.one ultra sp OPERATOR MANUAL
WARNING: Do not expose the external case of the LCD Display to any of the
following agents:
• Cidex
•Betadine
General
SmartCart/
SmartCart sp
Exterior
Surface
Cleaning
■
Do not use disinfectants (such as gluteraldahyde or hydrogen peroxide) or
acetone to clean any surfaces on the
SmartCart/SmartCart sp or their
accessories.
■
Before cleaning, turn off the AC circuit breaker on the SmartCart/SmartCart sp
to remove all power from the unit, and remove the battery from the
Scan
Engine
.
■
Wipe the SmartCart/SmartCart sp surfaces with a safe disinfectant solution
such as Sani-Cloth Plus or isopropyl 50% alcohol and follow the disinfectant
label instructions for use.
■
Do not autoclave, immerse, or attempt to sterilize the LCD display or
transducers.
■
Do not spill or spray liquid directly on the control panel, LCD display, battery
charger, AC power adapter, or transducer connector.
■
Using soap and water or a mild disinfectant, gently wipe the surfaces of the
SmartCart/SmartCart sp with a moistened cloth.
■
After each use, remove and dispose of any use cover/sheath. Wipe off any
excess gel from the transducer and clean it properly.
■
Air-dry the SmartCart/SmartCart sp.
WARNING: To avoid electrical shock before cleaning the SmartCart/SmartCart
sp
, turn off the AC circuit breaker at the rear of the SmartCart/SmartCart sp and
remove the battery from the
Scan Engine. Always use protective eyewear and
clothing when cleaning or disinfecting the
SmartCart/SmartCart sp.
■
The level of disinfection required for the SmartCart/SmartCart sp is dictated by
the type of exposure the
SmartCart/SmartCart sp came in contact with.
■
If premixed disinfection solution is used, observe the solution expiration date to
ensure that the solution has not expired.
CAUTION: Do not allow the disinfectant to contact metal surfaces. Use a soft cloth
and warm soapy water to remove any disinfectant that remains on metal surfaces.
Transducer Maintenance
NOTE: This section describes maintenance procedures for all ZONARE
transducers except the P8-3
TEE transducer. For the P8-3TEE transducer, see
page 6-14.
Regularly check transducers for signs of wear or damage.