7
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-71
Elastography Measurements
To Access Measurements for Elastography
1. Freeze the desired Elastography image.
2. Press Measure button. The Measurement options will display on the Imaging
menu.
Figure 7-34. Elastography Measurement Options
The following measurements are possible while using Elastography mode:
Distance Allows for linear measurement of the two-dimensional image. The Distance
measurement displays the span between two calipers on the image.
For instructions to do Distance measurements, see “Distance” on page 9-15. In
Elastography mode, the Distance measurements show simultaneously on both
images (B-Mode and Elastography).
Area (trace) Enables area measurement (cm
2
) and comparison of strain areas on the Dual
images (B-Mode and Elastography) displayed. The trace is free-form. The ratio of
the measured B-Mode area to the measured Elastography area is displayed in
lower right of screen.
Area (ellipse) This option is exactly like Area (trace) except the trace is always an ellipse.
Edge ✱
SmartCart/
SmartCart sp
■ Press Menu button and select desired Edge setting: E0, E1, E2, or E3.
NOTE: Edge allows for more or less spatial smoothing of display.
Table 7-7. Elastography User Controls (Continued)
Control SmartCart & SmartCart sp Scan Engine