7
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-51
Triplex Mode Triplex mode allows the user to visualize B-Mode, Color or Power Doppler, and
PW Doppler modes simultaneously.
• As with Duplex mode, an available Update option pauses either the 2D image
display (B and Color) or the scrolling PW strip.
•The Update ON/OFF control is the same as for Duplex mode.
Auto-Dop
Trace
The Auto-Dop Trace feature provides automatic tracing of waveforms on a PW
Doppler strip.
IMPORTANT: Only a basic overview of Auto-Dop Trace is provided in this chapter.
For detailed information on setup and operation, see
“Auto-Dop Trace” on
page 9-44 of this manual.
To Activate Auto-Dop Trace
1. While live scanning in any imaging mode, including PW Doppler, press
Function/Mode key assigned to Autotrace (see
page 9-46).
OR
If in PW Doppler mode, you can double-click the Measure button.
2. The system enters PW Doppler Autotrace mode and displays the Autotrace
Measure menu (
Figure 7-20). Autotrace icon appears next to PW in Doppler
Data Field (Figure 7-21; see also Figure 10-15, “Location of Doppler Data
Field,” on page 10-13).
NOTE: If you selected calculations on Auto-Dop Trace Configuration screen
(
page 9-45) as the default, the Autotrace Calc menu displays instead of Measure
menu.
Figure 7-20. AutoTrace Measure Menu
Select desired option and
press Set. Autotrace will
begin.